Salvage of Failed Total Ankle Replacement Using a Custom Titanium  Truss

Publication date: July–August 2016 Source:The Journal of Foot and Ankle Surgery, Volume 55, Issue 4 Author(s): Jennifer L. Mulhern, Nicole M. Protzman, Amari M. White, Stephen A. Brigido Subsidence of the talar component results in significant morbidity after total ankle replacement. When recognized, prompt revision could be needed to preserve the function of the implant; however, this is not always the case. In situations in which the implant cannot be revised, tibiotalocalcaneal arthrodesis might be necessary to salvage the extremity. The purpose of the present report is to describe the use of a custom titanium alloy truss to fill a bony void created by explantation of the implant components. Total ankle replacement was performed as the initial surgery to address end-stage osteoarthritis. Two years after the index procedure, the patient underwent revision of the polyethylene and talar components with subtalar arthrodesis secondary to progressive subtalar osteoarthritis and talar subsidence. The implant subsequently became infected and was removed. The patient underwent re-implantation after the infection had resolved, but significant talar subsidence required conversion to a tibiotalocalcaneal arthrodesis with a custom titanium alloy truss and retrograde intramedullary nail. At the most recent follow-up appointment, the patient was weightbearing on a stable extremity and pain free. Radiographic examination confirmed appropriate implant alignment and evidence o...