Estimation of the Dose of Radiation Received by Patient and Physician During a Videofluoroscopic Swallowing Study

In this study, we investigated the dose of radiation (entrance skin dose, ESD) estimated to be received by a patient during VFSS using a human phantom (via a skin-dose monitor sensor placed on the neck of the human phantom). We also investigated the effective dose (ED) and dose equivalent (DE) received by a physician (wearing two personal dosimeters) during an actual patient procedure. One dosimeter (whole body) was worn under a lead apron at the chest, and the other (specially placed to measure doses received by the lens of the eye) outside the lead apron on the neck collar to monitor radiation doses in parts of the body not protected by the lead apron. The ESD for the patient was 7.8 mGy in 5 min. We estimated the average patient dose at 12.79 mGy per VFSS procedure. The physician ED and DE during VFSS were 0.9 mSv/year and 2.3 mSv/year, respectively. The dose of radiation received by the physician in this study was lower than regulatory dose limits. However, in accordance with the principle that radiation exposure should be as low as reasonably achievable, every effort should be made (e.g., wearing lead glasses) to reduce exposure doses.
Source: Dysphagia - Category: Speech Therapy Source Type: research