On the accuracy of dose prediction near metal fixation devices for spine SBRT.

In this study, two commercially available dose prediction algorithms (Eclipse AAA and ACUROS), EGSnrc Monte Carlo, and GAFchromic film measurements were com-pared for a clinical spine SBRT case to determine their accuracy. An open six-field plan and a clinical nine-field IMRT plan were applied to a phantom containing a metal spine fixation device. Dose difference and gamma analysis were performed in and around the tumor region adjacent to the fixation device. Dose calculation inconsistency was observed in the open field plan. However, in the IMRT plan, the dose perturbation effect was not observed beyond 5 mm. Our results suggest that the dose effect of the metal fixation device to the spinal cord and the tumor volume is not observable, and all dose calculation algorithms evaluated can provide clinically acceptable accuracy in the case of spinal SBRT, with the tolerance of 95% for gamma criteria of 3%/3 mm. PMID: 27167252 [PubMed - in process]