Improved setup and positioning accuracy using a three-point customized cushion/mask/bite-block immobilization system for stereotactic reirradiation of head and neck cancer.

Improved setup and positioning accuracy using a three-point customized cushion/mask/bite-block immobilization system for stereotactic reirradiation of head and neck cancer. J Appl Clin Med Phys. 2016;17(3):6038 Authors: Wang H, Wang C, Tung S, Dimmitt AW, Wong PF, Edson MA, Garden AS, Rosenthal DI, Fuller CD, Gunn GB, Takiar V, Wang XA, Luo D, Yang JN, Wong J, Phan J Abstract The purpose of this study was to investigate the setup and positioning uncertainty of a custom cushion/mask/bite-block (CMB) immobilization system and determine PTV margin for image-guided head and neck stereotactic ablative radiotherapy (HN-SABR). We analyzed 105 treatment sessions among 21 patients treated with HN-SABR for recurrent head and neck cancers using a custom CMB immobilization system. Initial patient setup was performed using the ExacTrac infrared (IR) tracking system and initial setup errors were based on comparison of ExacTrac IR tracking system to corrected online ExacTrac X-rays images registered to treatment plans. Residual setup errors were determined using repeat verification X-ray. The online ExacTrac corrections were compared to cone-beam CT (CBCT) before treatment to assess agreement. Intrafractional positioning errors were determined using prebeam X-rays. The systematic and random errors were analyzed. The initial translational setup errors were -0.8 ± 1.3 mm, -0.8 ± 1.6 mm, and 0.3 ± 1.9 mm in AP, CC, and LR directions, respectively, ...