Anticoagulation, delivered dose and outcomes in CRRT: The program to improve care in acute renal disease (PICARD)

Abstract Delivered dialysis dose by continuous renal replacement therapies (CRRT) depends on circuit efficacy, which is influenced in part by the anticoagulation strategy. We evaluated the association of anticoagulation strategy used on solute clearance efficacy, circuit longevity, bleeding complications, and mortality. We analyzed data from 1740 sessions 24 h in length among 244 critically ill patients, with at least 48 h on CRRT. Regional citrate, heparin, or saline flushes was variably used to prevent or attenuate filter clotting. We calculated delivered dose using the standardized Kt/Vurea. We monitored filter efficacy by calculating effluent urea nitrogen/blood urea nitrogen ratios. Filter longevity was significantly higher with citrate (median 48, interquartile range [IQR] 20.3–75.0 hours) than with heparin (5.9, IQR 8.5–27.0 hours) or no anticoagulation (17.5, IQR 9.5–32 hours, P < 0.0001). Delivered dose was highest in treatments where citrate was employed. Bleeding complications were similar across the three groups (P = 0.25). Compared with no anticoagulation, odds of death was higher with the heparin use (odds ratio [OR] 1.82, 95% confidence interval [CI] 1.02–3.32; P = 0.033), but not with citrate (OR 1.02 95% CI 0.54–1.96; P = 0.53). Relative to heparin or no anticoagulation, the use of regional citrate for anticoagulation in CRRT was associated with significantly prolonged filter life and increased filter efficacy with respect to ...
Source: Hemodialysis International - Category: Hematology Authors: Tags: Original Article Source Type: research