Can we rely on machines? Device-detected atrial high rates correspond well with atrial arrhythmias in cardiac resynchronization recipients

Conclusion Nearly two-thirds of CRT-D patients had AHR episodes within 2.5 years after implantation. Almost 90% of AHRs were indeed AA. Misdetections were caused by far-field sensing or noise. A two-step diagnostic algorithm (>9% of time spent in AHRs and episode duration >36 s) allowed for proper detection of AA with a high hit-rate and specificity.
Source: Europace - Category: Cardiology Authors: Tags: Pacing and resynchronization therapy Source Type: research