Varian HDR surface applicators — commissioning and clinical implementation

The purpose of this study was to validate the dosimetric performance of Varian surface applicators with the source vertically positioned and develop procedures for clinical implementation. The Varian surface applicators with the source verti­cally positioned provide a wide range of apertures making them clinically advanta­geous, though the steep dose gradient in the region of 3–4 mm prescription depth presents multiple challenges. The following commissioning tests were performed: 1) verification of functional integrity and physical dimensions; and 2) dosimetric measurements to validate data provided by Varian as well as data obtained using the Acuros algorithm for heterogeneity corrected dose calculation. A solid water (SW) phantom was scanned and the Acuros algorithm was used to compute the dose at 5 mm depth and at surface for all applicators. Two sets of reference dose measurements were performed, with the source positioned at (i) -10 mm and (ii) -15 mm from the center of the first nominal dwell position. Measurements were taken at 5 mm depth in a SW phantom and in air at the applicator surface. The results were then compared to the vendor’s data and to the Acuros calculated dose. Relative dose measurements using Gafchromic films were taken at a depth of 4 mm in SW. Percent depth ionization (PDI) measurements using ion chamber were performed in SW. The profiles generated from film measurements and the PDI plots were compared with those computed using the Acuros algor...
Source: Journal of Applied Clinical Medical Physics - Category: Physics Source Type: research