Inappropriate Implantable Cardioverter-Defibrillator Shock from QRS Double Counting in the Setting of Hyperkalemia

Publication date: Available online 13 January 2016 Source:Cardiac Electrophysiology Clinics Author(s): Akash Dadlani, Jonathan W. Dukes, Nitish BadhwarTeaser This case shows the complexity of arrhythmia management in patients with implantable cardioverter-defibrillators (ICDs) who present with hyperkalemia. In order to prevent inappropriate ICD shock, consideration should be given to the suspension of ICD therapies while intensive care treatment of extreme electrolyte derangements is being pursued. Patients in these setting should be closely monitored until their electrocardiograms have normalized, after which the device can safely be reactivated.