Quantitation of S-Adenosylmethionine and S-Adenosylhomocysteine in Plasma Using Liquid Chromatography-Electrospray Tandem Mass Spectrometry

We describe a simple stable isotope dilution method for accurate determination of S-adenosylmethionine (SAM) and S-adenosylhomocysteine (SAH) in plasma as a diagnostic test. SAM and SAH are key metabolic intermediates of methionine metabolism and the methylation cycle. Determination of SAM and SAH in plasma was performed by high performance liquid chromatography coupled with electrospray positive ionization tandem mass spectrometry (HPLC-ESI-MS/MS). Calibrators (SAM and SAH) and internal standards (2H3-SAM and 2H4-SAH) were included in each analytical run for calibration. Sample preparation involved combining 20 μL sample with 180 μL of internal standard solution consisting of heavy isotope labeled internal standards in mobile phase A and filtering by ultracentrifugation through a 10 kd MW cutoff membrane. Sample filtrate (3 μL) was injected by a Shimadzu Nexera LC System interfaced with a 5500 QTRAP® (AB Sciex). Chromatographic separation was achieved on a 250 mm × 2.0 mm EA:faast column from Phenomenex. Samples were eluted at a flow rate of 0.20 mL/min with a binary gradient with a total run time of 10 min. The source operated in positive ion mode at an ion spray voltage of +5000 V. SAM and SAH resolved by a gradient to 100 % methanol with retention times of 6.0 and 5.7 min, respectively. The observed m/z values of the fragment ions were m/z 399 → 250 for SAM, m/z 385 → 136 for SAH, m/z 402 →...