Stent fracture and longitudinal compression detected on coronary CT angiography in the first- and new-generation drug-eluting stents

Abstract To evaluated prevalence and clinical implication of stent fracture and longitudinal compression in first- and new-generation drug-eluting stents (DES) using coronary computed tomography angiography (CCTA). The incidence of stent fracture and longitudinal compression were compared between first- and new-generation DES in 374 patients who underwent coronary stenting using DES and follow-up CCTA due to recurrent angina. 235 and 139 patients received 322 first- and 213 new-generation DES, respectively. The crude per-stent incidence of longitudinal compression (6.1 vs. 0.3 %, p < 0.001) was higher after new- than first-generation DES implantation using CCTA and the incidence of stent fracture (11.3 vs. 8.1 %, p = 0.23) was comparable. On follow-up coronary angiography for 347 stents, stent fracture (3.2 %) and longitudinal compression (0.9 %) were less detected than those on CCTA. Ostial stenting was a risk factor of longitudinal compression (p < 0.001). Stent fracture was associated with younger patients (p = 0.03), longer stent (p = 0.010), and excessively tortuous lesions (p = 0.001). The presence of stent fracture or longitudinal compression was not associated with poor clinical outcomes. The longitudinal compression more frequently occurred after new-generation DES implantation. The stent fracture was comparable between two DES. However, the occurrence of such mechanical deformities did not translate into a poor clinical outcom...
Source: The International Journal of Cardiovascular Imaging - Category: Radiology Source Type: research