Comparison of out-of-field normal tissue dose estimates for pencil beam scanning proton therapy: MCNP6, PHITS, and TOPAS

Biomed Phys Eng Express. 2022 Dec 23;9(1). doi: 10.1088/2057-1976/acaab1.ABSTRACTMonte Carlo (MC) methods are considered the gold-standard approach to dose estimation for normal tissues outside the treatment field (out-of-field) in proton therapy. However, the physics of secondary particle production from high-energy protons are uncertain, particularly for secondary neutrons, due to challenges in performing accurate measurements. Instead, various physics models have been developed over the years to reenact these high-energy interactions based on theory. It should thus be acknowledged that MC users must currently accept some unknown uncertainties in out-of-field dose estimates. In the present study, we compared three MC codes (MCNP6, PHITS, and TOPAS) and their available physics models to investigate the variation in out-of-field normal tissue dosimetry for pencil beam scanning proton therapy patients. Total yield and double-differential (energy and angle) production of two major secondary particles, neutrons and gammas, were determined through irradiation of a water phantom at six proton energies (80, 90, 100, 110, 150, and 200 MeV). Out-of-field normal tissue doses were estimated for intracranial irradiations of 1-, 5-, and 15-year-old patients using whole-body computational phantoms. Notably, the total dose estimates for each out-of-field organ varied by approximately 25% across the three codes, independent of its distance from the treatment volume. Dose discrepancies among...