Investigation the effect of a magnetic field on the dose distribution of I-125, Ir-192, Yb-169, and Co-60 brachytherapy sources by Monte Carlo simulation

This study investigates the impact of a magnetic field of 1.5 T, 3 T, and 7 T strengths on the dose distribution of high dose rate Co-60, Ir-192, and Yb-169, and low dose rate I-125 sources, using Geant4 Monte Carlo toolkit. After validating the simulation results by calculating the AAPM-TG43 dosimetric parameters, seven sources of each radioisotope were simulated in a water phantom, and their dose distributions were compared under the influence of a magnetic field. The simulation results indicate that using Co-60 brachytherapy under the MRI guidance is not recommended. Furthermore, the impact of a magnetic field of up to 7 T strength on the dose distribution of Ir-192, Yb-169, and I-125 sources is negligible, provided that there is no air pocket near brachytherapy sources.PMID:35717903 | DOI:10.1016/j.apradiso.2022.110332